Application Note : 005
Title of application :
Determination of bromate in water
Area of application :
Water Analysis
Conditions :
| Buffer | CE-AN001 Kit |
| Sample | 1. Bromate (10 ppb) in a water sample |
| Capillary | 30mm ID, 30 cm length |
| Injection mode | Electrokinetic injection at -7kV for 5 seconds |
| Separation setting | -7.2kV |
| Instrument used | CE-P1 |
Description :
Using electrokinetic injection , ppb level of bromate can be detected using CE-P1.
Determination of low levels of bromate anions are of interest due to their possible
carcinogenic properties. Bromate appears as possible by-product of the ozonation of
bromide-rich water (such as drinking water supplies from open sources near the sea).
Currently, the practical quantitation level for bromate in drinking water set by the US
EPA is 10mg/l (ppm) based on IC.