Application Note : 013

Title of application :
Detection of Ketamine as a Drug of Abuse from urine sample

Area of application :
Drug analysis/Biomedical analysis

Conditions :

Buffer KET-01 (20 mM), pH= 5.0;
Sample Urine sample containing 1 mg/L ketamine;
Capillary 50 mm i.d, 70 cm length;
Injection mode Hydrodynamic injection;
Separation setting + 20 kV
Instrument used CE-L1, UV detection at 210 nm

Description :
Detection of ketamine as an illicit drug from body fluids was commonly conducted using GC, GC/MS, LC or LC/MS. Apart from involving bulky and costly instrumentation, chromatographic analysis may also be complicated by the presence of interference due to the endogenous components in the sample matrix. In contrast, under the given CZE separation condition, ketamine was measured as the only component ahead of EOF, with all the endogenous compouds migrating near EOF because of their electrical neutrality. The excellent separation power, coupled with its operation simplicity, make CE-L1 a preferred alternative in the determination of ketamine and other illicit drugs. It is especially useful for drug regulatory agencies and law enforcement bodies as a routine means of drug screening.

pdf.gif (1283 bytes)Download application sheet in PDF format


Copyrights 2000 CE Resources Pte Ltd.  All rights reserved.