Application Note : 014
Title of application :
Detection of a recombinant interferon in an injectable dosage form
Area of application :
Biomedical analysis
Conditions :
| Buffer | CE-IF Kit |
| Sample | 1. Interferon a-2a; 2. Human serum albumin; 3. N-acetyltryptophan. |
| Capillary | 25 mm i.d, 40 cm length; |
| Injection mode | Pressure injection; |
| Separation setting | + 15kV |
| Instrument used | CE-L1, UV detection at 200 nm |
Description :
Interferon in injectable dosage forms can be detected by CE rapidly and sensitively.
Because of the high separation efficiency of CE, interferon can be well separated from
stabilizers, such as human serum albumin and N-acetyltryptophan.
Download application sheet in PDF format