| Light Source | Xenon lamp, 150 W |
| Wavelength Range | 200-650 nm (200-900nm option) |
| Wavelength Accuracy |
± 2nm |
| Wavelength Reproducibility |
± 0.2nm |
| Sensitivity | 300 of greater signal-to-noise ratio for Raman line of water |
| Cell Volume, Pressure |
On-columns detection cell |
| Time Programming | Excitation / emission wavelength range, response switching, wavelength scanning, auto zero, etc. |
| Dimensions | 260W X 520D X 205H mm |
| Weight | 17kg |
| Ambient Temperature Range |
4°C - 35 °C |
| Power Requirements |
100V, 350VA, 50/60Hz 120V, 350VA, 50/60Hz 220-240V, 350VA, 50/60Hz |
| Noise Specification |
Delivers high sensitivity detection of ± 0.35 x 10-5 AU. |
| Baseline Stability |
Isolation of the light source compartment and of the monochromator reduces temperature variation, minimizing drift and reducing baseline noise. |
| Wavelength Reproducibility |
The wavelength reproducibility (± 0.1 nm) given high reliability. Stability is guaranteed with a wavelength accuracy specification of ± 1 nm. |
| Wavelength Measurement Range |
Wavelength range analysis 190 - 600 nm. |
| Simultaneous Dual-Wavelength Measurement |
Can acquire data at two different wavelengths in a single analytical run, gives qualitative information on peak purity, improving analysis reliability. |
| Wavelength Scanning Capability |
The wavelength scanning function subtracts mobile phone noise in the background spectrum so that truer spectrums are attained. |
| Deuterium Lamp | The new deuterium lamp lasts twice as long as older versions (~2000 hrs). |
| Light Sources | D2 lamp, W lamp, or both (3 modes) |
| Wavelength Range |
190 - 800 nm |
| Wavelength Accuracy |
1nm |
| Noise Level | 0.8 x 10-5 AU (250 nm, 600 nm) |
| Drift | < 1 x 10-3 AU/hr (250 nm, 600 nm) |
| Cell | 10 mm; 50 kgf/cm2 (4.9 MPa); SUS 316, Quartz, PTFE (optical pathlength, pressure, wetted surfaces) |
| Dimensions: | 260 mm W x 420 mm D x 140 mm H |
| Weight: | 10.5kg |
| Ambient Temperature |
4°C - 35°C |
| Power Requirments |
100, 120, 220-240 V as ordered, 150 VA, 50/60 Hz |
| Measuring ranges: | 0-2,0-20, 0-200 RFU |
| Dynamic range: | 105, linear within 2% |
| Sensitivity: | S/N>10 for a solution of 10-12 M of Rhodamine in a flow cell capillary of 320 m m ID with a 488 nm Ar-Ion laser 10 mW. |
| Noise Level: | 0.003 RFU flow cell capillary of 320 mm ID with a 488 nm Ar-Ion laser 10 mW. |
| Dimension and weight: | 247 x 370 x 217 mm (W x D x H), 8 kg. |
| Ambient temperature range: | 10-35 °C. |
| Power requirements: | 220-240 V, 0.5 A, 50/60 Hz or 100-120 V, 1 A, 50/60 Hz for the detector. Lasers available at the same voltage ratings, current from 1 to 10 A, according to the laser power. |
| Outputs: | 1 V (unprocessed). 10 mV, 100 mV and 1 V (processed). |
| Lasers: | Ar-Ion at 488 nm and 514 nm, and He-Cd at 325 nm and 442 nm. Any other wavelength on request. |
| Flow cell capillaries: | I..D. 25, 50, ., 320 mm. |
| Dimension: | 225 x 525 x 250 (W x D x H) |
| Ambient temperature range: | 10-35 °C. |
| Power requirements: | 220-240 V, 50/60 Hz, 50 W |
| Outputs: | -10V to +10V |
**Contents of this page are subjected to change without prior notice.