Fluorescence Detector

Ultra-Violet (UV) Detector

Photodiode Array Detector

Laser Induced Fluorescence Detector

Potential Gradient Detector (CE-D1)

 

Fluorescence Detector

Light Source Xenon lamp, 150 W
Wavelength Range 200-650 nm (200-900nm option)
Wavelength
Accuracy
± 2nm
Wavelength
Reproducibility
± 0.2nm
Sensitivity 300 of greater signal-to-noise ratio for Raman line of water
Cell Volume,
Pressure
On-columns detection cell
Time Programming Excitation / emission wavelength range, response switching, wavelength scanning, auto zero, etc.
Dimensions 260W X 520D X 205H mm
Weight 17kg
Ambient
Temperature Range
4°C - 35 °C
Power
Requirements
100V, 350VA, 50/60Hz
120V, 350VA, 50/60Hz
220-240V, 350VA, 50/60Hz

Back To Top...

Ultra-Violet (UV) Detector

Noise
Specification 
Delivers high sensitivity detection of ± 0.35 x 10-5 AU.
Baseline 
Stability 
Isolation of the light source compartment and of the monochromator reduces temperature variation, minimizing drift and reducing baseline noise.
Wavelength
Reproducibility 
The  wavelength reproducibility (± 0.1 nm) given high reliability. Stability is guaranteed with a wavelength accuracy specification of ± 1 nm.
Wavelength
Measurement
Range
Wavelength range analysis 190 - 600 nm.
Simultaneous
Dual-Wavelength
Measurement
Can acquire data at two different wavelengths in a single analytical run, gives qualitative information on peak purity, improving analysis reliability.
Wavelength
Scanning
Capability
The wavelength scanning function subtracts mobile phone noise in the background spectrum so that truer spectrums are attained.
Deuterium Lamp The new deuterium lamp lasts twice as long as older versions (~2000 hrs).

Back To Top...

Photodiode Array Detector

Light Sources D2 lamp, W lamp, or both (3 modes)
Wavelength
Range
190 - 800 nm
Wavelength
Accuracy
1nm
Noise Level 0.8 x 10-5 AU (250 nm, 600 nm)
Drift < 1 x 10-3 AU/hr (250 nm, 600 nm)
Cell 10 mm; 50 kgf/cm2 (4.9 MPa); SUS 316, Quartz, PTFE (optical pathlength, pressure, wetted surfaces)
Dimensions: 260 mm W x 420 mm D x 140 mm H
Weight: 10.5kg
Ambient
Temperature
4°C - 35°C
Power
Requirments
100, 120, 220-240 V as ordered, 150 VA, 50/60 Hz

Back To Top...

Laser Induced Fluorescence Detector

Measuring ranges: 0-2,0-20, 0-200 RFU
Dynamic range: 105, linear within 2%
Sensitivity: S/N>10 for a solution of 10-12 M of Rhodamine in a flow cell capillary of 320 m m ID with a 488 nm Ar-Ion laser 10 mW.
Noise Level: 0.003 RFU flow cell capillary of 320 mm ID with a 488 nm Ar-Ion laser 10 mW.
Dimension and weight: 247 x 370 x 217 mm (W x D x H), 8 kg.
Ambient temperature range: 10-35 °C.
Power requirements: 220-240 V, 0.5 A, 50/60 Hz or 100-120 V, 1 A, 50/60 Hz for the detector.  Lasers available at the same voltage ratings, current from 1 to 10 A, according to the laser power.
Outputs: 1 V (unprocessed).  10 mV, 100 mV and 1 V (processed).
Lasers: Ar-Ion at 488 nm and 514 nm, and He-Cd at 325 nm and 442 nm. Any other wavelength on request.
Flow cell capillaries: I..D. 25, 50,……., 320 mm.

Back To Top...

 

Potential Gradient Detector (CE-D1)

Dimension: 225 x 525 x 250 (W x D x H)
Ambient temperature range: 10-35 °C.
Power requirements: 220-240 V, 50/60 Hz, 50 W
Outputs: -10V to +10V

Back To Top...

**Contents of this page are subjected to change without prior notice.